Merge Healthcare

84 recall(s) on record — most severe on record: Class II.

Recalls by year

FDA actions on record by year2011: 2 recalls; 2012: 2 recalls; 2013: 1 recalls; 2014: 3 recalls; 2015: 14 recalls; 2016: 44 recalls; 2017: 9 recalls; 2018: 1 recalls; 2019: 1 recalls; 2020: 2 recalls; 2021: 1 recalls; 2025: 2 recalls; 2026: 2 recalls2’112’121’133’1414’1544’169’171’181’192’201’21’22’23’242’252’26
  • Recalls (84)

Registered facilities (2)

Recalls & enforcement

20262 records
  • Class IIOngoing2026-07-22
    Software issue may cause measurements to be displayed inaccurately.
  • Class IIOngoing2026-06-30
    Thumbnails from a previously opened patient study may be displayed in the Thumbnail Bar while a different patient study is currently open.
20252 records
  • Class IIOngoing2025-08-11
    Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain units trigger an error message on the Merge Hemo Client (Pressure Sensor Defective) during non-invasive blood pressure (NIBP…
  • Class IIOngoing2025-02-19
    When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements i…
20211 record
  • Class IITerminated2021-07-23
    The firm identified a safety issue affecting the software associated with Merge Hemo versions 9.x and 10.x, used in combination with an HL7 lab interface.
20202 records
  • Class IITerminated2020-08-06
    Measurements done on the Merge PACS generated MPR s may have incorrect measurements.
  • Class IITerminated2020-05-08
    An error in the calculation of the Michigan Reference Ranges can generate an inaccurate Z-Score calculation. The error may reduce the sensitivity of the Z-Score in detecting an abnormality resulting in a false negative.
20191 record
  • Class IITerminated2019-07-11
    Potential for prior thumbnails to not display in reverse chronological order and images may not scroll in reverse chronological order when mammography stacked scrolling is enabled. Merge PACS provides image manipulation tools to enable users to view and compare images such a…
20181 record
  • Class IITerminated2018-12-20
    Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) an…
20179 records
  • Class IITerminated2017-08-14
    The software is not identifying the patient as having atypical hyperplasia, resulting in an incorrect Gail Risk calculation.
  • Class IITerminated2017-08-07
    The software is not identifying the patient as having atypical hyperplasia, resulting in an incorrect Gail Risk calculation.
  • Class IITerminated2017-06-13
    When the user has not set up any user preference on the sorting order to render the study images (OS/OD/etc.) and reports, the ECP may get an incorrect study index. If the user attempts to delete an image, ECP may then read that image as OS when it is, in fact, OD.
  • Class IITerminated2017-05-24
    A migrated study that has annotations will display the annotations, but when a prior study is loaded that should have annotations, the second study will not display annotations.
  • Class IITerminated2017-05-24
    Studies that have QC, PDE, or HL7 ADT updates applied may fail to propagate via telmed protocol to the destination server and may fail to import on the destination server.
  • Class IITerminated2017-05-08
    Merge PACS did not show unviewed images when the last view was skipped. There is a potential risk to health of a physician misdiagnosis because not all images are available for viewing.
  • Class IITerminated2017-05-08
    When cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis.
  • Class IITerminated2017-04-06
    Merge is conducting a hardware recall (replacement) of Fellows "Mighty 8" Power Strip and TrippLite Power Strip due to overheat of input resistor causing the power strip to smoke.
  • Class IITerminated2017-01-23
    A defect in the software resulted in medications that are not associated with the patient (i.e., medications that the patient is not currently taking) appearing on their report.
201644 records
  • Class IITerminated2016-12-09
    This recall has been initiated due to an issue related to the potential accidental deletion of record(s) by an Eye Station user.
  • Class IITerminated2016-12-09
    Merge received reports of Merge Eye Station being unable to capture images of the eye as expected per the intended use of the product. Merge is recalling product from the field to reduce the risk to patient health.
  • Class IITerminated2016-10-20
    A product issues happens for RF projection images. If user measures on RF image, .cal (Calibration) is not displayed and measurement is not correct. The .cal label is not displayed on the projection images after calibration is performed which should be present if the measurement …
  • Class IITerminated2016-07-14
    It was reported by a customer that RadSuite images are not appearing as readable images. This product malfunction has the potential to result in delay in care and a possible re-exposure of the patient to radiation (x-ray).
  • Class IITerminated2016-05-25
    Software displayed incorrect prior reports in the viewport area, only when more than one prior study (2 or more) was viewed.
  • Class IITerminated2016-04-15
    There are potential issues with results reporting for certain run-based tests. Under certain conditions, the wrong results could inadvertently be verified.
  • Class IITerminated2016-04-13
    Fetal patient report was automatically pulling prior measurement data for a prior fetus since the fetal study is based on the mother's MRN, resulting in the wrong fetal measurements getting referenced.
  • Class IITerminated2016-04-08
    Reports were missing demographic header information when they were faxed to referring physicians.
  • Class IITerminated2016-04-08
    The software is not adding the correct tomo image.
  • Class IITerminated2016-04-04
    A situation can occur allowing two physicians to access the same study report in EchoIMS when launched from the Cardio Study List without receiving the read-only notification prompt.
  • Class IITerminated2016-04-04
    Cardio study list does not show STAT studies without refreshing.
  • Class IITerminated2016-04-04
    The versions allow images to be stored without pixel data, resulting in the system not being able to present all the prior studies, which could cause a delay in treatment in determining the progression of disease.
  • Class IITerminated2016-04-04
    The system will completely stop allowing the downloading or acquisition of new images due to incompatible software application versions.
  • Class IITerminated2016-04-04
    The study is archived but cannot be opened in iConnect Access and cannot send to PACS resulting in comparison studies not being available for physician review to determine the progression of disease.
  • Class IITerminated2016-04-04
    System locks up which may result in potential patient injury or delay in diagnosis or treatment.
  • Class IITerminated2016-04-04
    Studies that have QC, PDE, or HL7 ADT updates applied may fail to propagate via telmed protocol to the destination server and may fail to import on the destination server.
  • Class IITerminated2016-04-04
    When taking measurements from images on the Cardio workstation or from the US cart, numbers are not crossing to the report in the correct unit of measure.
  • Class IITerminated2016-04-04
    Under certain conditions, the cables used with the Hemo system during a patient procedure could stop displaying the patient SpO2 causing a slight delay in patient care.
  • Class IITerminated2016-04-04
    During an antivirus program scan of the Eye Station or Eye Care PACS, the antivirus program may detect a newly captured image to be an unwanted file and delete the image permanently.
  • Class IITerminated2016-04-04
    Use of the software may show an incorrect value to the user when viewing the Fractional Flow Reserve (FFR) results during recording.
  • Class IITerminated2016-04-04
    The software produced a number of "do not route" exceptions, which may result in potential patient injury or delay in diagnosis or treatment.
  • Class IITerminated2016-04-04
    In some instances, the system will lock tabs within a study, even when a second user does not have the study open on a different workstation.
  • Class IITerminated2016-04-04
    Potential exists for an incorrect patient image being displayed which could result in the delay in diagnosis or treatment.
  • Class IITerminated2016-04-04
    An error message can occur resulting in the prior studies being unavailable for viewing.
  • Class IITerminated2016-03-28
    If a reader selects Multi-study review prior to the original study completing its loading process and then immediately returns to image review of the original study, it is possible that not all current images will be present for review, which could result in incorrect treatment/d…
  • Class IITerminated2016-03-28
    If images are sent without an order in the system, they will be matched with the latest order on the current patient/modality matching potentially resulting in the matching of the report to the incorrect accession number (but still associated to the correct patient).
  • Class IITerminated2016-03-28
    Eye Station images were not importing properly and were imported under "unknown" due to an issue when validating patients using only an Medical Record Number (MRN)
  • Class IITerminated2016-03-28
    Cut lines on the image may present horizontally rather than vertically.
  • Class IITerminated2016-03-28
    The software did not show unviewed images when the last view was skipped. The physician may potentially misdiagnose when not all images are available for viewing.
  • Class IITerminated2016-03-28
    There is a potential connection issue when powering up the Merge Hemo Record Station and the Hemo Monitor does not communicate with the Client PC.
  • Class IITerminated2016-03-10
    If the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the selected pressure waveform being provided by the PDM to the Link Assembly which provides the analog output.
  • Class IITerminated2016-02-09
    A report can be confirmed with the incorrect patient demographics resulting in it being saved under the incorrect patient record in the Electronic Medical Record (EMR).
  • Class IITerminated2016-02-05
    When measuring a lesion on an unmagnified mammography image then performing the same measurement on an image magnified by the Hologic imager (different image), the measurements are not the same.
  • Class IITerminated2016-01-31
    Potential data loss occurs as a result of product archiving not working properly.
  • Class IITerminated2016-01-31
    Potential for RadSuite AV viewer to skip image slices when the user presses page up or page down when the viewer displays multiple viewports and the user has mouse focused on any viewport other than the first viewport.
  • Class IITerminated2016-01-30
    Potential data loss occurs as a result of the software archiving not working properly.
  • Class IITerminated2016-01-30
    Patient information in the header is only found on the first page of the report and not on the subsequent pages.
  • Class IITerminated2016-01-30
    Studies coming over via telmed were missing patient's DOB, procedure, and referring physician.
  • Class IITerminated2016-01-30
    Communication protocols interfacing with the affected software version with some select instruments were not properly handled, preventing data captured by these instructions to be processed by the rest of the LIS software.
  • Class IITerminated2016-01-30
    A migrated study that has annotations will display the annotations, but when a prior study is loaded that should have annotations, the second study will not display annotations.
  • Class IITerminated2016-01-30
    The patient name in the Halo title bar and the thumbnails do not match the name on displayed images.
  • Class IITerminated2016-01-30
    Interventional Radiology (IR) images are stored as JPEG2k Lossless in Merge Enterprise Archive (EA) and are not displaying correctly in RadSuite.
  • Class IITerminated2016-01-30
    The software was not presenting the PURE filter user interface message on DICOM transfer studies. If a site starts using PURE filter, the user interface message will not display. This may result in incorrect patient follow-up.
  • Class IITerminated2016-01-30
    Studies that are viewed in iConnect Access that originate on Merge PACS / Ortho PACS may not display the correct current patient demographics if there has been a PDE (Patient / Study Demographic Edit) performed after the images were ingested in PACS.
201514 records
  • Class IITerminated2015-12-17
    Potential incorrect Standardized Uptake Values (SUV) measurements in RadSuite.
  • Class IITerminated2015-11-17
    The values provided from the Pixel Value tool do not appear to be correct, which may result in potential patient injury or delay in diagnosis or treatment.
  • Class IITerminated2015-10-26
    The firm, Merge Healthcare, sent out Merge HEMO V10.0 & Merge CARDIO V10.1 software to 3 customers before the product was completely validated in house.
  • Class IITerminated2015-10-15
    The software fails to associate to the correct MG image if there are two images for the same view.
  • Class IITerminated2015-10-08
    Measurements in the software are changing after saving a plan with a measurement result, which may result in potential patient injury or delay in diagnosis or treatment.
  • Class IITerminated2015-10-01
    Reporting feature times out after inactivity for more than an hour sending the user back to the study list, which causes all reporting data being entered to be lost.
  • Class IITerminated2015-09-22
    Users can merge a device import file with an image study that already has a confirmed report, which may result in including information not present at the time of physician interpretation.
  • Class IITerminated2015-09-22
    For sites using the Issuer of Patient ID (IPID), the system will display the study list and images from different patients with different IPIDs as though they are for the same patient if they all have the first name, last name, and medical record number in common.
  • Class IITerminated2015-07-23
    The software is unable to accurately determine the calcium score of scans with a slice thickness not equal to 3 mm.
  • Class IITerminated2015-05-31
    Non-invasive blood pressure pumps (NIBP) voltage is not sufficient and causes the NIBP pumps to fail prior to reaching adequate initial pressure.
  • Class IITerminated2015-04-09
    When RadSuite is used with IPID (Issuer of Patient ID) as a part of the "Patient Identifier," it is possible in some circumstances that the demographics of one patient will be applied to a study or studies for another patient.
  • Class IITerminated2015-03-30
    The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential of a small bias added to the CO2 measurement and affect the alarm notification.
  • Class IITerminated2015-03-05
    There is a potential for duplicate container numbers to be created for patients.
  • Class IITerminated2015-01-21
    A system freeze-up of the Merge Hemo system, that included the PHASEIN End Tidal CO2 unit, resulted in the user needing to move the patient to another cath lab.
20143 records
  • Class IITerminated2014-09-29
    It has been reported that during use, the SpO2 value displayed on the Hemo Monitor may not update to reflect changes in the patient's oxygen value. It is also possible that if using the pulse rate of the SpO2 finger clip to calculate the patient's heart rate, it too may be subje…
  • Class IITerminated2014-08-28
    There is an issue related to eFilm Workstation versions 4.0 and 4.0.1 when having multiple studies open and utilizing the thumbnail panel to select multiple series from multiple studies may lead to the selection of an incorrect study.
  • Class IITerminated2014-04-30
    The application may crash during the cath lab procedure.
20131 record
20122 records
  • Class IITerminated2012-11-01
    An incorrect biopsy or missed target could result if the incorrect grid is selected within the application.
  • Class IITerminated2012-07-02
    The International Normalized Ratio (INR) value displayed in the study report does not match the value that is imported on the pre-procedure labs screen from HL7
20112 records
  • Class IITerminated2011-10-28
    After a period of time running Fusion Workstation, the Hounsfield measurement tool will report incorrect values.
  • Class IITerminated2011-10-19
    Customers may experience an issue with the software study preferences when changes are made to the study protocol, resulting in incorrect patient follow-up.

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.

Cite this record

RegWarden. “Merge Healthcare — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-10024020-merge-healthcare/

Not open yet

Alerts for Merge Healthcare

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.