Spiegelberg GmbH & Co. KG
1 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (1)
Registered facilities (1)
- SPIEGELBERG GmbH & Co. KG FEI 3002108226 · DE
Recalls & enforcement
- Class IITerminated2009-12-29In the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were displayed in the lower measuring range from 0 to 20 mmHg. The observed measurement error was particularly at the lowest end o…
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “Spiegelberg GmbH & Co. KG — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-10037562-spiegelberg-gmbh-co-kg/
Alerts for Spiegelberg GmbH & Co. KG
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.