  • Class IIOngoing2025-11-06
    Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.
  • Class IIOngoing2023-09-26
    Affected lots of product may experience technical distortions to a greater extent than expected.