Mirion Technologies, Inc.
3 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (3)
Registered facilities (2)
- MIRION TECHNOLOGIES (CAPINTEC), INC FEI 2518443 · NJ
- Mirion Technologies (Capintec), Inc. FEI 3042471748 · CT
Recalls & enforcement
- Class IIOngoing2026-07-20Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.
- Class IIOngoing2024-04-22Complaints of unexpected detachment of the collimator have been reported.
- Class IIOngoing2023-11-27Spring arm failure with Captus 4000e Thyroid Uptake Systems, can cause the collimator to fall downward to its lowest point of travel, causing potential injury with a patient or operator
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “Mirion Technologies, Inc. — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-10043529-mirion-technologies-inc/
Alerts for Mirion Technologies, Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.