Contego Medical, INC
2 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (2)
Registered facilities (2)
- CONTEGO MEDICAL, INC FEI 3025465918 · CA
- CONTEGO MEDICAL, INC FEI 3011471056 · NC
Recalls & enforcement
- Class IIOngoing2026-06-29Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss…
- Class IIOngoing2025-09-04Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter that may not open or may not fully open during sheath retraction and operation of the filter knob. The stent is used with a primary distal embolic protection device.
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “Contego Medical, INC — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-10050012-contego-medical-inc/
Alerts for Contego Medical, INC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.