SEDECAL S.A.
11 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (11)
Registered facilities (4)
- A.T.S. Applicazione Tecnologie Speciali s.r.l. FEI 3011347852 · IT
- SEDECAL S.A. FEI 3002495042 · ES
- Stephanix FEI 3006972752 · FR
- Metaltronica Spa FEI 3009868110 · IT
Recalls & enforcement
- Class IIOngoing2025-06-10Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- Class IIOngoing2024-01-22Potential interference with the anti-fall system.
- Class IIOngoing2023-12-13Some parts have fallen off the equipment resulting in a safety risk to patients and/or users.
- Class IIOngoing2023-05-23There is a software login in issue that may prevent the user from logging in.
- Class IIOngoing2023-02-06Sedecal was informed of 3 incidents of erratic movements related to X-ray mobile systems.
- Class IIOngoing2022-08-12Due to un-commanded movement of mobile x-ray unit.
- Class IIOngoing2021-10-14While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on i…
- Class IIOngoing2018-07-18A safety problem with installation and maintenance could lead to the breakage of the two steel cables that support the equipment to its roof anchor. Due to this, the equipment may fall and cause harm to the patient, user, or third parties.
- Class IITerminated2016-01-18This is related to previous 806 Notifications about uncommanded movements of a motorized mobile x-ray system. Reason for Recall: Along with the previously noted encoder and gauges problems and corresponding fixes, we have observed that static electricity can also elicit the unwan…
- Class IITerminated2015-07-13Electrostatic energy may be stored in Systems that can cause a short-cicuit which, in turn, can generate an Electomagnetic Pulse which might disturb the electronics of the Digital Motion Control Board producing uncontrolled movements and such uncontrolled movements could cause th…
- Class IITerminated2014-08-06Reports of unexpected movement.
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “SEDECAL S.A. — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-9039772-sedecal-s-a/