CENORIN

4 recall(s) on record — most severe on record: Class II.

Recalls by year

FDA actions on record by year2013: 1 recalls; 2014: 1 recalls; 2017: 1 recalls; 2018: 1 recalls1’131’14’15’161’171’18’19’20’21’22’23’24’25’26
  • Recalls (4)

Registered facilities (1)

Recalls & enforcement

  • Class IITerminated2018-04-09
    There is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn.
  • Class IITerminated2017-06-07
    Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn.
  • Class IITerminated2014-04-24
    Cenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential risk of the system not injecting the proper amount of cleaning fluid during the wash cycle and not achieving the desired level of debris reduction.
  • Class IITerminated2013-01-24
    The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.

Cite this record

RegWarden. “CENORIN — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-9057829-cenorin/

Not open yet

Alerts for CENORIN

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.