RegWardenFDA facility compliance records
Facility

MEDICA CORP.

4 recall(s) on record — most severe on record: Class II.

Activity by year

FDA actions on record by year2017: 2 recalls, 1 inspections; 2019: 1 recalls; 2025: 1 recalls3201720181201920202021202220232024120252026
  • Recalls (4)
  • Inspections (1)

Devices registered here (3)

Recalls & enforcement

  • Class IIOngoing2025-12-31
    A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.
  • Class IITerminated2019-02-06
    Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.
  • Class IITerminated2017-11-21
    The firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an ineffective wash solution resulting in an interference carryover…
  • Class IITerminated2017-04-11
    Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.

FDA inspections (1)

  • NAI2017-04-14
    No Action Indicated · Compliance: Devices

Source: U.S. FDA Inspections Dashboard (opens in a new tab) (device inspection classifications, 2016–present). Data as of 2026-10-06. The dashboard profile can take several seconds to load.

Related

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)

Cite this record

RegWarden. “MEDICA CORP. (FEI 1220972) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/1220972-medica-corp/

Not open yet

Alerts for MEDICA CORP.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.