THE DAAVLIN DISTRIBUTING CO.
3 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (3)
Devices registered here (1)
- Light, Ultraviolet, DermatologicalProduct code FTC (opens in a new tab) · Class II · General, Plastic Surgery · 21 CFR 878.4630
Recalls & enforcement
- Class IIOngoing2026-03-05Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Class IIOngoing2025-12-10Some of the Calibration/Output Certificates had the outputs reversed.
- Class IIOngoing2022-04-26Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under dosing.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “THE DAAVLIN DISTRIBUTING CO. (FEI 1526255) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/1526255-the-daavlin-distributing-co/