Facility

CIVCO Medical Instruments Co., Inc.

11 recall(s) on record — most severe on record: Class II.

Activity by year

FDA actions on record by year2012: 1 recalls; 2013: 1 recalls; 2014: 1 recalls; 2015: 2 recalls; 2016: 1 inspections; 2018: 3 recalls; 2021: 1 recalls; 2025: 1 recalls; 2026: 1 recalls1’121’131’142’151’16’173’18’19’201’21’22’23’241’251’26
  • Recalls (11)
  • Inspections (1)

Devices registered here (1)

Recalls & enforcement

  • Class IIOngoing2026-03-02
    There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
  • Class IIOngoing2025-05-01
    During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.
  • Class IITerminated2021-10-08
    There is potential that the VirtuTRAX device will not tighten onto the instrument to be used.
  • Class IITerminated2018-11-02
    The needle guide bracket may puncture the sterile cover applied over the transducer and bracket during assembly.
  • Class IITerminated2018-07-30
    The assembly of the needle guides were found to be either missing adhesive or did not have adequate adhesive to hold the assembled cannula in place on the needle guide.
  • Class IITerminated2018-02-23
    A cover failed leak testing during validation testing of new needle guides from a replacement tool.
  • Class IITerminated2015-10-21
    Sterility of the product cannot be assured.
  • Class IITerminated2015-02-26
    Leaks were detected in the recalled lot which may compromise the sterility of the product.
  • Class IITerminated2014-08-01
    A pinhole leak identified in the packaging of the recalled product potentially could compromise the sterility of the product.
  • Class IITerminated2013-10-25
    Scales adhered to the product may be misaligned from the zero reference such that one side will not match the other.
  • Class IITerminated2012-05-07
    The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.

FDA inspections (1)

  • NAI2016-06-30
    No Action Indicated · Compliance: Devices

Source: U.S. FDA Inspections Dashboard (opens in a new tab) (device inspection classifications, 2016–present). Data as of 2026-10-06. The dashboard profile can take several seconds to load.

Related

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)

Cite this record

RegWarden. “CIVCO Medical Instruments Co., Inc. (FEI 1937223) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/1937223-civco-medical-instruments-co-inc/