EBI PATIENT CARE, INC.
10 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (10)
- Inspections (2)
Devices registered here (1)
- Stimulator, Bone Growth, Non-InvasiveProduct code LOF (opens in a new tab) · Class II · Orthopedic · 21 CFR 890.5870
Recalls & enforcement
- Class IIOngoing2024-03-22Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment
- Class IITerminated2014-11-11A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation.
- Class IITerminated2014-10-08Inoperative screw head seat splay and/or fracture as a result of positioning of the rod which is a part of the Cypher MIS Screw System.
- Class IITerminated2014-05-15The Cypher MIS Screw Inserter may exhibit an increased rate of instrument tip failure.
- Class IITerminated2013-09-24Biomet received a complaint on 5/20/2013 reporting that the tip of the instrument sheared off during a removal case.
- Class IITerminated2013-08-22Polaris Translation Screw Failures were recently reported to Biomet Spine. These failures are generally described as dissociation of the screw head from the shaft of the Polaris Translation Screw.
- Class IITerminated2013-08-13Several complaints were reported to exhibit screw head splay in the Cypher MIS Screw System.
- Class IITerminated2013-04-09Several complaints were recorded regarding one of the three screws driving through the cage of the Solitaire 35 implant.
- Class IITerminated2013-02-07Increased risk of failure at the driver tip during screw insertion or removal.
- Class IITerminated2012-09-29Broken screw insertor hardness was measured and found to be below the acceptable limits.
FDA inspections (2)
- NAI2019-03-12No Action Indicated · Compliance: Devices
- VAI2018-01-17Voluntary Action Indicated · Compliance: Devices · citations posted
Source: U.S. FDA Inspections Dashboard (opens in a new tab) (device inspection classifications, 2016–present). Data as of 2026-10-06. The dashboard profile can take several seconds to load.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “EBI PATIENT CARE, INC. (FEI 3003435252) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3003435252-ebi-patient-care-inc/
Alerts for EBI PATIENT CARE, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.