NORDICNEUROLAB AS
8 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (8)
Devices registered here (1)
- System, Image Processing, RadiologicalProduct code LLZ (opens in a new tab) · Class II · Radiology · 21 CFR 892.2050
Recalls & enforcement
- Class IITerminated2022-06-16When exporting merged results from the BOLD, DSC, and DCE modules of nordicBrainEx, where both the underlay and overlay originated from a multiframe dataset, the resulting output may have left/right sides flipped.
- Class IITerminated2018-08-20Complaints of this version of the power supply unit (LPS) (1.0 and 1.1) being dragged into the magnet bore (MR Scanner).
- Class IITerminated2017-10-28Report was of an interrupted MR scan after patient complained about acoustic distortion, and heat felt form the headphones.
- Class IITerminated2015-07-03The bug causes BOLD activation maps to be visualized as overlays without taking the coregistration into account. Any output created from these activation maps as overlays will not be adjusted according to the coregistration.
- Class IITerminated2014-12-15An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
- Class IITerminated2014-09-17An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
- Class IITerminated2014-09-12An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
- Class IITerminated2012-12-21There is an error related to the relative geometry between fiber tracts in a fiber group and images. The error occurs in the following situations: A group of fibers has been selected using the VOI functionality and put into a fiber group. This is done with images series A visu…
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “NORDICNEUROLAB AS (FEI 3006738448) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3006738448-nordicneurolab-as/
Alerts for NORDICNEUROLAB AS
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.