VIEWRAY SYSTEMS, INC.
6 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (6)
Devices registered here (2)
- System, Nuclear Magnetic Resonance ImagingProduct code LNH (opens in a new tab) · Class II · Radiology · 21 CFR 892.1000
- Accelerator, Linear, MedicalProduct code IYE (opens in a new tab) · Class II · Radiology · 21 CFR 892.5050
Recalls & enforcement
- Class IITerminated2016-03-07When editing the isocenter or the couch position of the plan while in the treatment workflow (in the Points screen) and re-optimizing, the software will not prompt the user to shift the couch to the new isocenter. As a result there is the potential to deliver dose to the initia…
- Class IITerminated2015-12-24The system loaded a completion fraction in the incorrect order after a treatment interruption.
- Class IITerminated2015-08-27ViewRay discovered that in the event that an encoder breaks or fails on the Patient Handling System (PHS, or couch), when attempting to restart the system, the couch could move unexpectedly.
- Class IITerminated2015-05-07The software was failing to determine new patient locations if imaging is not enabled during treatment.
- Class IITerminated2015-04-01ViewRay received a report that the couch moved unexpectedly into the bore after performing a RTCS reboot.
- Class IITerminated2015-01-15The software was not correctly using the RT (Radiation Therapy) to MR (Magnetic Resonance image) coordinate correction for non HFS (Head First Supine) patient orientations, resulting in slice mismatch error.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “VIEWRAY SYSTEMS, INC. (FEI 3007546534) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3007546534-viewray-systems-inc/
Alerts for VIEWRAY SYSTEMS, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.