Synaptive Medical Inc
4 recall(s) on record — most severe on record: Class I.
Activity by year
- Recalls (4)
- Inspections (1)
Devices registered here (1)
- Neurological Stereotaxic InstrumentProduct code HAW (opens in a new tab) · Class II · Neurology · 21 CFR 882.4560
Recalls & enforcement
- Class IITerminated2021-04-23Due to visible burrs/metal filings affixed along the suction tube inner perimeter.
- Class IITerminated2020-02-26A software defect negatively affects the quality of images taken during diagnostic imaging (mammograms) which may result in misdiagnosis.
- Class ITerminated2018-10-09This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the software defect is triggered…
- Class IITerminated2015-12-04Out of tolerance for radio frequency emissions. At the 150-1000MHz frequency, the testing indicated the BrightMatter Navigation system was up to 20dB uV/meter higher than the applicable IEC 60601-1-2:2007 (Ed3.0) standard specification.
FDA inspections (1)
- VAI2025-01-23Voluntary Action Indicated · Compliance: Devices · citations posted
Source: U.S. FDA Inspections Dashboard (opens in a new tab) (device inspection classifications, 2016–present). Data as of 2026-10-06. The dashboard profile can take several seconds to load.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Synaptive Medical Inc (FEI 3012075008) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3012075008-synaptive-medical-inc/
Alerts for Synaptive Medical Inc
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.