WW-West
1 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (1)
Devices registered here (1)
- Component, External, Limb, Ankle/FootProduct code ISH (opens in a new tab) · Class I · Physical Medicine · 21 CFR 890.3420
Recalls & enforcement
- Class IIOngoing2021-04-20the firm discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body, causing the cord to come loose or slip which could result in the loss of suspension.
Related
- All facilities operated by WillowWood Global LLC
- WillowWood Global LLC (OH)
- Xtremity by WillowWood Global (CO)
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “WW-West (FEI 3035765256) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3035765256-ww-west/
Alerts for WW-West
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?