RegWardenFDA facility compliance records
Recall

Class II — Pride Mobility Products Corp

Recalling firmPRIDE MOBILITY PRODUCTS CORP.
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe main harness connector was manufactured without the required back shield sealing protection. As a result, there is the potential for water to seep through the nonconforming harness.
ProductVictory 9 with Power Seat (SC609PS) Motorized wheelchair with elevating seat.
Quantity186
DistributionWorldwide Distribution
Initiated2010-11-26
Classified2012-07-03
Terminated2013-03-22
Recall numberZ-1941-2012
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)