Recall

Class I — Baxter Healthcare Corporation

Recalling firmBAXTER HEALTHCARE CORPORATION
Registered facilityView this company's 32 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusTerminated
ReasonRecall expansion; the infusion pumps have the potential to fail causing inaccurate flow conditions during use. These conditions range from back flow to free flow, which could result in over-infusion. Sigma's evaluation of subsequent complaints involving failed bearings indicated that the loss or degradation of bearing lubricant to be a more significant contributor to bearing failure than initially believed. Based upon the results of the analysis of these additional failures, Sigma determined that any bearings prior to the latest, improved design may be at risk for failure. Sigma decided to expand the recall of the Spectrum Infusion Pumps to include units manufactured from 1/18/2005 through 11/01/2010.
ProductSigma Spectrum VOLUMETRIC INFUSION PUMP, with or without Master Drug Library, Part Number 35700 -- Common/Classification Name: Infusion Pump -- Class 2 medical device -- SIGMA LLC, Medina, NY 14103 Product Usage: The Spectrum Infusion Pump is intended to be used for the controlled administration of intravenous fluids. These fluids may include blood, blood products or mixtures of pharmaceutical drugs for required patient therapy. The Spectrum is used in conjunction with legally marked intravenous administration sets and medications provided by the user. The Master Drug Library is a softw…
Quantity63,225 pumps total (corrected number): Domestic - 63,032 pumps; Canada - 29 pumps; Okinawa - 164 pumps
DistributionWorldwide Distribution - USA Nationwide including Puerto Rico and the countries of Canada and Okinawa
Initiated2011-07-06
Classified2012-06-13
Terminated2017-10-17
Recall numberZ-1490-2012
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Baxter Healthcare Corporation (Z-1490-2012).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/61355/

Not open yet

Alerts for BAXTER HEALTHCARE CORPORATION

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.