RegWardenFDA facility compliance records
Recall

Class I — Natus Medical Incorporated

Recalling firmNatus Medical Incorporated
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusTerminated
ReasonThe Olympic Cool-Cap's Control Module has experienced a frozen screen during use. When this occurs, the on-screen information remains on display, but the system is no longer providing cooling treatment to the infants. The visible clock displayed at the upper right hand corner of the screen stops advancing.
ProductThe product has the Catalogue/Part number 60010. Olympic Cool-Cap System is a selective head cooling system for treatment of hypoxic-ischemic encephalopathy (HIE) in infants. A cooling water cap is placed on the infant's head and the body is warmed using radiant warmers. The Cool-Cap provides 72 hours of monitored cooling. The device is labeled in parts: "***OLYMPIC COOL-CAP Cooling Unit***Natus Medical Incorporated 5900 First Avenue South ***Made in USA***Control Unit Model No. 60010***Software in this product is***Technical Service***"
Quantity324 units
DistributionWorldwide distribution: USA (nationwide) including states of: AK, AR, AZ, CA, CO, DE, FL, GA, HI, ID, IL, IN, KY, MA, MI, MN, MS, ND, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, WA, Puerto Rico, and countries of: Argentina, Australia, Austria, Bahrain, Canada, Chile, China, Colombia, Ecuador, Egypt, Germany, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Italy, Kuwait, Latvia, Malaysia, M…
Initiated2012-05-09
Classified2012-11-30
Terminated2015-01-27
Recall numberZ-0447-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Natus Medical Incorporated (Z-0447-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/61835/