Class II — Abbott Laboratories

Recalling firmABBOTT LABORATORIES
Registered facilityView this company's 64 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe six ARCHITECT CA 19-9XR reagent lots are demonstrating a shift up in patient results. The Abbott Controls however do not detect the upward shift and will test within range.
ProductARCHITECT CA 19-9XR Reagent Kit; ARCHITECT CA 19-9 Reagent Kit, list 291-20 (4 x 100 tests), 2K91-25 and 2K91-27 (1 x 100 tests); Product Usage: The ARCHITECT CA 19-9XR assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of 1116-NS-19-9 reactive determinants in human serum or plasma on the ARCHITECT i System. The ARCHITECT CA-19-9XR assay is to be used as an aid in the management of pancreatic cancer patients in conjunction with other clinical methods.
Quantity316 kits - USA, 8,356 kits ROW
DistributionWorldwide Distribution - US (nationwide) the states of: Alabama, Arkansas, Arizona, California, Florida, Iowa, Indiana, Louisiana, Michigan, Minnesota, New Jersey, Nevada, New York, Ohio, Oklahoma, Pennsylvania, South Dakota, Texas, Virginia, Washington, including Puerto Rico, and the countries of Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Barbados, Belarus, Belgium, Bosnia and …
Initiated2012-05-30
Classified2012-06-28
Terminated2016-12-23
Recall numberZ-1904-2012
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Abbott Laboratories (Z-1904-2012).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/61979/

Not open yet

Alerts for ABBOTT LABORATORIES

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.