Class II — Sekisui Diagnostics Llc

Recalling firmSekisui Diagnostics, LLC
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonDiscoloration of the acetaminophen enzyme reagent (R1), generates a high calibration factor and erratic control recovery.
ProductMULTIGENT Acetaminophen for use with Architect/Aeroset Cat. No. 2K99-20. Assay is intended for the quantitative determination of acetaminophen in human serum or plasma.
Quantity1600 kits
DistributionIllinois
Initiated2012-05-29
Classified2012-06-27
Terminated2012-11-14
Recall numberZ-1902-2012
InitiationVoluntary: Firm initiated