Class II — Terumo Cardiovascular Systems Corporation
| Recalling firm | Terumo Cardiovascular Systems Corp. |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav |
| Product | Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The safety monitor is indicated to warn when the blood level in the reservoir reaches the operator-set limit, and to stop the arterial and cardioplegia pumps. |
| Quantity | 8 |
| Distribution | Worldwide distribution: International only-country of: Terumo Dubai, UAE |
| Initiated | 2009-08-12 |
| Classified | 2012-09-11 |
| Terminated | 2013-01-29 |
| Recall number | Z-2348-2012 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Terumo Cardiovascular Systems Corporation (Z-2348-2012).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/62321/
Alerts for Terumo Cardiovascular Systems Corp.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.