Class II — Belmont Instrument Corporation

Recalling firmBELMONT INSTRUMENT LLC.
Registered facilityBELMONT MEDICAL TECHNOLOGIES (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonLeak from a saline bag caused saline to drip along the back of the Rapid infuser close to the Power Entry Module causing an electrical flash at the power entry area.
ProductBelmont Rapid Infuser, a Fluid Management System, Model FMS2000.
Quantity2,913 units
DistributionWorldwide Distribution including Nationwide (USA) and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Denmark, England, Estonia, France, Germany, Hong Kong, India, Israel, Italy, Jordan, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Peru, Puerto Rico, S. Korea, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, UK, U…
Initiated2012-09-25
Classified2012-10-12
Terminated2014-02-28
Recall numberZ-0058-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Belmont Instrument Corporation (Z-0058-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/62703/

Not open yet

Alerts for BELMONT INSTRUMENT LLC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.