RegWardenFDA facility compliance records
Recall

Class II — DePuy Spine, Inc.

Recalling firmDePuy Spine, Inc.
Registered facilityDEPUY SPINE, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonA component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification.
ProductDePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct.
Quantity140 units
DistributionWorldwide Distribution - USA (nationwide) and internationally to Belgium, Egypt, Ireland, Italy, Portugal, and the UK.
Initiated2012-07-24
Classified2013-01-16
Terminated2013-12-30
Recall numberZ-0685-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — DePuy Spine, Inc. (Z-0685-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/62719/