Class II — Integra LifeSciences Corporation

Recalling firmIntegra LifeSciences Corporation
Registered facilityView this company's 15 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonT-handle screw is used on a complementary product to the CRW, the Integra HRAIM, which is the CT-compatible intubation Head Ring Assembly. When the T-handle on the HRAIM intubation hoop is tightened, the T-handle could stop in a vertical position that prevents the CRWPRECISE and certain models of the CRW-ASL from seating properly on the head ring although it appears to be fully seated
ProductIntegra HRAIM Head Ring Assembly with Intubation Mounts The CRW stereotactic system is used in conjunction with the HRAIM head ring is a target-centered system that can be configured to be CT-only or CT/MR compatible. The CRW stereotactic system is a multi-purpose system used for localizing intercranial targets for precisely directing instruments such as: Biopsy forceps RF lesioning electrodes Deep brain electrodes Recording and stimulating electrodes
Quantity188 units
DistributionWorldwide distribution: USA (Nationwide) and countries of: Argentina, Australia, Belgium, Canada, China, Czec, Denmark, France, Hong Kong, Hungary, Italy, Japan, Malaysia, Philippines, Singapore, Spain, South Africa, and UK.
Initiated2012-08-09
Classified2012-09-20
Terminated2014-03-26
Recall numberZ-2429-2012
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Integra LifeSciences Corporation (Z-2429-2012).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/62978/

Not open yet

Alerts for Integra LifeSciences Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.