Class II — Encore Medical, Lp
| Recalling firm | ENCORE MEDICAL, L.P. |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | A product complaint was received where a size 20 screw was labeled as size 30. After receipt, measurement of the screw determined it to be a size 25. Further investigation also determined that there were other screws labeled as size 25, but a size 30 screw was inside of the package. It was thought that all of these screws were found in-house and were contained, but it was determined later that on |
| Product | 6.5mm Cancellous Bone Screw Intended to be used for the fixation of the acetabular shell or fracture repair. |
| Quantity | 1 unit |
| Distribution | Nationwide Distribution only to CA. |
| Initiated | 2012-09-28 |
| Classified | 2012-10-25 |
| Terminated | 2013-01-25 |
| Recall number | Z-0139-2013 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Encore Medical, Lp (Z-0139-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/63321/
Alerts for ENCORE MEDICAL, L.P.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.