Class II — Arobella Medical, LLC
| Recalling firm | Arobella Medical, LLC |
|---|---|
| Registered facility | AROBELLA MEDICAL, LLC (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | This recall has been initiated because the Arobella Medical 6mm Qurette Q07 Lot Numbers 0001 and 0002 can fracture during use and may release a fragment or fragments. The 6 mm Qurette is a component of the AR1000 Quostic Wound Therapy System. |
| Product | Arobella Medical 6mm Qurette used with the Arobella Medical AR1000 Wound Therapy System. Major components of the AR1000 Wound Therapy System include an ultrasonic generator, converter hand piece, done-shaped applicator tip (qurette), and a saline supply reservoir. Product Usage: The Arobella Medical AR1000 Wound Therapy System is interned for the selective dissection and fragmentation of tissue, wound debridement ( of acute and chronic wounds, burns, and diseased or necrotic tissue), and cleansing irrigation of the wound site for the removal of debris, exudates, fragments, and other m… |
| Quantity | 99 |
| Distribution | Worldwide distribution - US (nationwide) and the states of AL, CA, ID, IN, IA, MI, MN, MS, NJ, NM, OH, PA. SC, TX, UT, VA, WI. including the country of SAUDI ARABIA. |
| Initiated | 2012-11-05 |
| Classified | 2012-12-17 |
| Terminated | 2015-03-04 |
| Recall number | Z-0534-2013 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Arobella Medical, LLC (Z-0534-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/63808/
Alerts for Arobella Medical, LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.