RegWardenFDA facility compliance records
Recall

Class II — Oxus America, Inc.

Recalling firmOxus America Inc
Registered facilityOxus America, Inc. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonDue to past battery inventory storage conditions and deficiencies in the maintenance procedure, it was possible that some batteries were over-discharged to point of damaging the cells.
ProductLi-Ion battery used in the Oxus Portable Oxygen Concentrator (POC), RS-00400. Also branded as Reliability Plus byDrive. The Portable Oxygen Concentrator is intended to deliver concentrated oxygen for adult patients with chronic pulmonary diseases.
Quantity1724 total units
DistributionWorldwide Distribution-USA including the states of New York, Michigan, Colorado, Texas, Iowa, Florida, Georgia, Ohio, Wisconsin, Oregon, Alabama, Illinois, Utah, Oklahoma, West Virginia and California and the countries of Canada, European Union, South Africa, Korea and Brazil.
Initiated2013-01-14
Classified2013-02-02
Terminated2013-07-17
Recall numberZ-0768-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Oxus America, Inc. (Z-0768-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/64057/

Not open yet

Alerts for Oxus America Inc

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.