Class II — Philips Medical Systems (Cleveland) Inc
| Recalling firm | Philips Medical Systems International BV |
|---|---|
| Registered facility | View this company's 54 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The four screws that secure the patient table top to the carrier became loose, which caused the table top to detach. |
| Product | Philips MX 16-Slice CT Scanner System, a whole-body computed tomography X-ray system. |
| Quantity | 412 Units |
| Distribution | Worldwide Distribution - USA including AL, AR, CA, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NV, NY, OH, PA, PR and VA. |
| Initiated | 2012-12-19 |
| Classified | 2013-02-28 |
| Terminated | 2013-12-03 |
| Recall number | Z-0899-2013 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)