RegWardenFDA facility compliance records
Recall

Class II — Intuitive Surgical, Inc.

Recalling firmIntuitive Surgical, Inc.
Registered facilityView this company's 41 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonIntuitive Surgical has identified a potential for some units of the Monopolar curved Scissors version 09 and 10 instruments to develop very small cracks near the distal end. This may create a pathway for electrosurgical energy to leak to tissue and potentially cause thermal injury.
Product8 mm Monopolar Curved Scissors (a.k.a. Hot Shears) Manufactured and Distributed by: Intuitive Surgical Sunny vale, CA The EndoWrist Monopolar Curved Scissors Instrument is a multiple-use endoscopic instrument utilizing a single use, tip cover accessory, to be used in conjunction with the Intuitive Surgical Endoscopic Instrument Control System.
Quantity~132,441 instruments
DistributionWorldwide distribution - USA (nationwide) Foreign Distribution to Argentina, Austria, Australia, Belgium, Brazil, Bulgaria, Canada, China, Chile, Czech Republic, Denmark, Finland, France, Germany, Cyprus- Greece, Egypt, Guadeloupe, Israel, Ireland, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Monaco, Netherlands, New Zealand, Norway, Pakistan, Philippines, Portugal, Poland, …
Initiated2013-05-08
Classified2013-05-31
Terminated2014-05-01
Recall numberZ-1442-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Intuitive Surgical, Inc. (Z-1442-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/65216/