Class II — Sysmex America Inc

Recalling firmSYSMEX CORPORATION
Registered facilityView this company's 14 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe data printed on the thermal printer may omit a digit or decimal point in the results or sample identification number.
ProductpocH-100i; sysmex; Automated Hematology Analyzer; 7.3" x 13.8" x 18.1" / 30.8 lbs.; Sysmex Corporation; 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073, Japan; Sysmex America, Inc; One Nelson C. White Pkwy, Mundeline, IL 60060, U.S.A. The Sysmex pocH-100i Automated Hematology Analyzer is an automated cell counter intended for in vitro diagnostic use in CLIA non-waived clinical laboratories (not for Point of Care use in a CLIA waived laboratory). The Complete Blood Cel (CBC) count test parameters include: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, and NEUT%/#, LYMPH %/…
Quantity193 units
DistributionWorldwide distribution: US (Nationwide) including Puerto Rico and states of: AK, AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, SD, TN, TX, UT, VA, VT, WA, WI, and WY; and country of Canada.
Initiated2013-05-15
Classified2013-10-29
Terminated2016-08-15
Recall numberZ-0100-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Sysmex America Inc (Z-0100-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/65722/

Not open yet

Alerts for SYSMEX CORPORATION

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.