Class II — Fresenius Medical Care Holdings, Inc.

Recalling firmFresenius Medical Care Holdings, Inc.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonUpdated Instructions for Use: Bloodline connections not tightened per instructions in the operator's manual may disconnect and lead to blood loss
ProductCombiset Blood Line Hemodialysis use Catalog Number: 03-2622-3. Intended for use as the extracorporeal blood circuit during Hemodialysis and is indicated for use with conventional and high flux negative pressure hemodialyzer equipment.
Quantity508,029 cases
DistributionWorldwide Distribution-USA (nationwide) and the country of Canada.
Initiated2013-08-29
Classified2013-11-08
Terminated2021-02-11
Recall numberZ-0186-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Fresenius Medical Care Holdings, Inc. (Z-0186-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/65739/

Not open yet

Alerts for Fresenius Medical Care Holdings, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.