Class II — B. Braun Medical, Inc.

Recalling firmB. BRAUN MEDICAL INC.
Registered facilityView this company's 5 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonReports were received of an incorrectly assembled Outlook Safety Infusion System Burette Set (Product Code 375038), and the investigation determined that one of the ULTRASITE injection sites and corresponding low pressure backcheck valve was assembled upside down.
ProductOutlook Safety Infusion System Burette Set. For intravenous fluid administration.
Quantity900
DistributionDistributed in the states of FL, NY, NJ, and TX.
Initiated2013-08-05
Classified2013-09-06
Terminated2014-09-03
Recall numberZ-2166-2013
InitiationVoluntary: Firm initiated