RegWardenFDA facility compliance records
Recall

Class II — Maquet Cardiovascular Us Sales, Llc

Recalling firmMaquet Cardiovascular LLC
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe manufacturer has conducted a review of the results of product testing performed during the last twelve months and has identified specific lots of product which may not meet specifications for bond strength on the inlet and/or outlet connector.
ProductQUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000. The QUADROX oxygenators are intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.
Quantity1,182 units
DistributionWorldwide Distribution, including Nationwide (US) and foreign countries.
Initiated2013-08-15
Classified2013-09-20
Terminated2017-07-11
Recall numberZ-2262-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Maquet Cardiovascular Us Sales, Llc (Z-2262-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/66049/