Recall

Class II — Aesculap, Inc.

Recalling firmAESCULAP, INC.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonComplaints were received for the GN161 Bipolar Foot Control reporting that the bipolar energy did not stop after release of the foot control.
ProductAesculap GNI161 Bipolar Foot Control Product Usage: The Aesculap Bipolar Coagulator is intended for use in surgery to generate electrical power for bipolar instruments. Bipolar coagulators are used in neurosurgery, ENT surgery, urology, laparoscopy, and plastic surgery.
Quantity85
DistributionWorldwide Distribution - USA Nationwide and countries of: Austria, Germany, Japan, and Taiwan.
Initiated2013-09-10
Classified2013-12-10
Terminated2014-10-20
Recall numberZ-0479-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Aesculap, Inc. (Z-0479-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/66533/

Not open yet

Alerts for AESCULAP, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.