Recall

Class II — ConMed Corporation

Recalling firmCONMED CORPORATION
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe affected lots were exposed to a higher than acceptable level of particulate matter during manufacture and subsequently distributed. ConMed evaluated the potential risk that this issue could present to the patient and concluded the risk of injury to be negligible; however exposure could cause a minor, local, self correcting infection.
ProductALTRUS® THERMAL TISSUE FUSION, Tissue Fusion Handpiece, 5 mm O.D., 36 cm Shaft, REF 60-9522-001, STERILE -- CONMED CORPORATION 525 FRENCH RD, UTICA, NY 13502 --- The ConMed Altrus® Thermal Tissue Fusion System is comprised of a dedicated energy source and disposable handpiece used to ligate (seal) and divide (cut) blood vessels and tissue bundles that fit into the jaws of the handpiece. The Altrus® System utilizes a thermal energy platform to achieve the desired clinical effect. The ConMed Altrus® Thermal Tissue Fusion handpiece is a single use device which is provided sterile. The device use…
Quantity147 handpieces (Domestic: 135 devices; Foreign: 12 devices)
DistributionWorldwide distribution: US (Nationwide) and countries of: Canada, Germany and Italy.
Initiated2013-10-23
Classified2013-11-15
Terminated2017-03-08
Recall numberZ-0332-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — ConMed Corporation (Z-0332-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/66547/

Not open yet

Alerts for CONMED CORPORATION

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.