RegWardenFDA facility compliance records
Recall

Class II — Urologix, Inc.

Recalling firmUrologix LLC
Registered facilityUROLOGIX, LLC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonA Hand Piece high deployment force exceeded the maximum specification.
ProductProstiva RF Therapy Hand Piece Kit, REF 8929, Rx only, sterilized using Ethylene Oxide. The Prostiva RF Model 8929 Hand Piece Kit is the delivery system component of the Prostiva RF Therapy System. Prostiva RF Therapy delivers low-levels of radiofrequency (RF) energy directly into the hyperplastic tissue for the treatment of benign prostatic hyperplasia (BPH). The RF energy delivered by the system provides selective thermal ablation of prostate tissue to help reduce urethral constriction and relieve BPH voiding symptoms. The Prostiva RF Therapy System consists of the following major compo…
Quantity142
DistributionNationwide Distribution-including the states of AZ, CA, FL, GA, IL, KY, MA, MN, NC, NJ, NV, NY, OH, OR, TN, TX, and UT.
Initiated2013-10-21
Classified2013-11-08
Terminated2014-04-21
Recall numberZ-0182-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Urologix, Inc. (Z-0182-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/66598/

Not open yet

Alerts for Urologix LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.