Class II — Integra LifeSciences Corp.
| Recalling firm | Integra LifeSciences Corporation |
|---|---|
| Registered facility | View this company's 15 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Integra became aware through an adverse complaint trend that Rickham Style Reservoirs with bent or broken connectors that were shipped in sleeve-type packages were found prior to use but after the sterile package was open. |
| Product | Integra Large Rickham -Style Reservoir Rx Only Sterile for Single Use Only Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Made in the USA NL 850-1132 Provides access to the lateral cerebral ventricles via hypodermic puncture. It is useful in obtaining CSF (cerebral spinal fluid_ samples for cytological and chemical studies, for monitoring ventricular fluid pressure and for ventricular drainage. The reservoir provides easy access to the lateral ventricles and to cystic tumors for the injections of chemotherapeutic agents and/or radio-isotopes. The rickham reservoir ma… |
| Quantity | 78 packages |
| Distribution | Worldwide Distribution - USA including FL, NE, WI, WA, CT, ND, AL, NY, NH, OR, OK, MN, and TX. Internationally to Belgium, Czech Republic, Germany, France, Great Britian, Italy, Poland, and Portugal. |
| Initiated | 2013-11-13 |
| Classified | 2014-01-03 |
| Terminated | 2015-02-13 |
| Recall number | Z-0622-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Integra LifeSciences Corp. (Z-0622-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/67025/
Alerts for Integra LifeSciences Corporation
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.