Class II — Integra LifeSciences Corp.

Recalling firmIntegra LifeSciences Corporation
Registered facilityView this company's 15 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonIntegra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is used on CAM01 monitor that contains Mod 4 Firmware, the displayed values for the systolic and diastolic intracranial pressure (ICP) are reversed. But, if the CAM01 is used with an auxiliary bedside monitor, the bedside monitor displays these values correctly.
ProductIntegra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, 311 Enterprise Dr Plainsboro, NJ 08536 USA The Camino Advanced Monitor is a compact, portable device for use with Camino 110-4 series of Pressure / Temperature and Pressure Transducer-Tipped Catheters. Pressure and/or Temperature are measured at the Catheter tip, eliminating the need for external transducers, fluid, pressure tubing, and flush devices. The Camino displays intracranial pressure (ICP), intracranial temperature (ICT), and calculates cerebral perfusion pressure (CPP) (CPP = mean arterial pre…
Quantity275 CAM01 Monitors
DistributionWorldwide Distribution - USA ( Nationwide ) and the countries of : UK, Saudi Arabia, South Africa, Brunei, China, France, Poland, Turkey, Greece, Sweden, United Arab Emirates, Russia, Spain, Pakistan, Finland, Israel, Italy, Slovenia, Japan, Poland, Malaysia, Germany and Panama.
Initiated2013-11-26
Classified2013-12-27
Terminated2015-02-13
Recall numberZ-0602-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Integra LifeSciences Corp. (Z-0602-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/67094/

Not open yet

Alerts for Integra LifeSciences Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.