RegWardenFDA facility compliance records
Recall

Class II — AGFA Corp.

Recalling firmAgfa N.V.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonPatient name displayed (printed) on the Patient Report was the wrong patient name.
ProductIMPAX RIS QDOC 5.8
Quantity14
DistributionCA, FL, GA, ME, OH, OR, SC, SD, TX and Canada
Initiated2014-01-08
Classified2014-02-03
Terminated2014-02-28
Recall numberZ-0919-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — AGFA Corp. (Z-0919-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/67275/

Not open yet

Alerts for Agfa N.V.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.