Class I — Draeger Medical Systems, Inc.

Recalling firmDraeger Medical Systems, Inc.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusTerminated
ReasonThe potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.
ProductDraeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833
Quantity3
DistributionUSA Nationwide Distribution in the states of WI and NY.
Initiated2014-03-05
Classified2014-04-10
Terminated2014-10-31
Recall numberZ-1440-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)