Recall

Class II — Philips Medical Systems North America Inc.

Recalling firmPhilips Medical Systems International BV
Registered facilityView this company's 54 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonA problem has been detected with Philips Model EPIQ 5 and 7 Ultrasound System. The system will report incorrect Doppler velocities if you use High Pulse Repetition Frequency (HPRF) with multiple sample volume gates. While in High Pulse Repetition Frequency (HPRF) in Display Zoom Velocity Scale using manual spectral Doppler trace, the velocities in the velocity results table may be overstated b
ProductPhilips Model EPIQ 7 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis. Catalogue numbers 795200 / 795201 / 795202 / 795204 / 795205 / 795206 EPIQ is intended for diagnostic ultrasound imaging and fluid flow analysis. The collection of Doppler velocity results values are used to assist in diagnosing the existence, location and severity of cardiac or vasculare stenosis.
Quantity1267 total :670 units in US and 1267 units Internationally (which includes 11 units in Canada)
DistributionWorldwide Distribution: US (nationwide) including states of: AL, AK, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NM, NJ, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, and WV; and countries of: Australia, Austria, Belgium, Canada, Chili, Ecuador, Egypt, Finland, France, Germany, Greece, India, Ireland, Italy, Japan, Korea Rep, Malaysia, Netherla…
Initiated2014-02-28
Classified2014-03-20
Terminated2016-12-12
Recall numberZ-1264-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Philips Medical Systems North America Inc. (Z-1264-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/67685/

Not open yet

Alerts for Philips Medical Systems International BV

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.