Class II — Siemens Healthcare Diagnostics, Inc.

Recalling firmSIEMENS HEALTHCARE DIAGNOSTICS INC.
Registered facilityView this company's 12 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonA small number of Vista Air Compressors were miswired, which if installed and powered up, could cause a fire, or become an electrical shock hazard.
ProductSiemens Healthcare Diagnostics Vista Air Compressor. Vista AirCompressor replacement part: SMN 10706557 SKIT Air Compressor Assy, used on the Dimension Vista 500 and Dimension Vista 1500 systems. Siemens Healthcare Diagnostics Dimension Vista System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. Dimension Vista chemical and immunochemical applications use photometric, turbidimetric, chemiluminescence, nephelometric and inte…
Quantity4
DistributionWorldwide Distribution: US Distribution to states of: IL, OH and TN; and country of: Italy.
Initiated2014-04-14
Classified2014-06-09
Terminated2014-12-15
Recall numberZ-1733-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Healthcare Diagnostics, Inc. (Z-1733-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68076/

Not open yet

Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.