Class II — The Anspach Effort, Inc.
| Recalling firm | THE ANSPACH EFFORT, INC. |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Anspach product code L-3SD contained product code S-3SD which is shorter in length than the labeled product. |
| Product | ANSPACH Cutting bur, 3 mm Diamond Ball, Standard Length. REF L-3SD Use with LONG Attachments. Used for cutting and shaping bone, including the spine and cranium. |
| Quantity | 30 cutting burs affected |
| Distribution | Nationwide Distribution including FL, TX, NC, IL, CO, and PA. |
| Initiated | 2014-01-08 |
| Classified | 2014-05-09 |
| Terminated | 2015-01-20 |
| Recall number | Z-1574-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — The Anspach Effort, Inc. (Z-1574-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68109/