Class II — Boston Scientific CRM Corp

Recalling firmBoston Scientific Corporation
Registered facilityView this company's 37 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonA review of manufacturing test records for Boston Scientific ENDOTAK RELIANCE implantable leads revealed a suspected test data recording error. Specifically, some test results were recorded as "failed" without any other indication of failure. There are no reported injuries from the devices.
ProductBoston Scientific ENDOTAK RELIANCE® SG, transvenous defibrillation lead, Models 0180 & 0292, Sterile EO. Product Usage: The ENDOTAK RELIANCE leads provide pacing and rate-sensing and deliver cardioversion and defibrillation shocks for automatic implantable cardioverter defibrillator (AICD) systems.
Quantity21
DistributionWorldwide Distribution -US (nationwide) in the states of OH, MN and country of France
Initiated2014-06-19
Classified2014-08-04
Terminated2014-12-03
Recall numberZ-2147-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Boston Scientific CRM Corp (Z-2147-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68700/

Not open yet

Alerts for Boston Scientific Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.