Class II — Siemens Medical Solutions USA, Inc
| Recalling firm | SIEMENS MEDICAL SOLUTIONS USA, INC. |
|---|---|
| Registered facility | View this company's 9 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Siemens has become aware of the potential malfunction when using syngo Imaging XS, version VA70A or higher. In some cases a miscalculation may occur for the grey scale values in the functions Region of Interest, Pixel Lens, Edge enhancement and Histograms. This potential malfunction may occur for images with a pixel depth greater than 12 bit. All values calculated for images with a pixel d |
| Product | Siemens syngo Imaging XS, version VA70A or higher Syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo Imaging XS also supports storage and archiving of Structured DICOM reports. Syngo Imaging XS optionally uses a variety of advanced post processing applications. |
| Quantity | 141 |
| Distribution | Nationwide Distribution |
| Initiated | 2014-06-30 |
| Classified | 2014-07-23 |
| Terminated | 2015-02-24 |
| Recall number | Z-2111-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc (Z-2111-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68769/
Alerts for SIEMENS MEDICAL SOLUTIONS USA, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.