Class II — Ion Beam Applications S.A.

Recalling firmION BEAM APPLICATIONS S.A.
Registered facilityION BEAM APPLICATIONS S.A. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonWhile the Proteus 235 is in the pencil beam scanning (PBS) treatment mode, on rare occasions, Low Level Radio Frequency (LLRF) pulsed dee voltage (i.e., voltage applied between two electrodes in the cyclotron to accelerate protons) may generate a hot spot on a pencil beam position within a tumor, generating a dose that might exceed the prescribed dose on a single voxel.
ProductProteus 235 Proton Therapy System for cancer treatment.
Quantity8
DistributionWorldwide Distribution-USA (nationwide) including the states of PA, NJ, IL, WA, and TN, and the countries of Germany, Italy, and Czech Republic.
Initiated2014-03-27
Classified2014-10-29
Terminated2014-12-10
Recall numberZ-0020-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Ion Beam Applications S.A. (Z-0020-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68846/

Not open yet

Alerts for ION BEAM APPLICATIONS S.A.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.