Class II — Ortho Development Corporation

Recalling firmOrtho Development Corporation
Registered facilityORTHO DEVELOPMENT CORP. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonOrtho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.
ProductPatella 38mm, REF164-0038 Product Usage: UHMWPE component used in a TKA procedure to resurface the patella and serve as a bearing surface for sliding contact with the femoral component
Quantity14
DistributionWorldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.
Initiated2014-07-11
Classified2014-08-17
Terminated2014-10-06
Recall numberZ-2254-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Ortho Development Corporation (Z-2254-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68898/

Not open yet

Alerts for Ortho Development Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.