Class II — Curbell Medical, Inc.
| Recalling firm | CURBELL MEDICAL PRODUCTS, INC. |
|---|---|
| Registered facility | CURBELL MEDICAL PRODUCTS, INC. (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The firm became aware of a potential problem that was initiated by a customer complaint. After consultation with the manufacturer, it was discovered that a resistor was incorrectly placed within the circuit board on the monitor. This change to the resistor was a planned change to address a product improvement (improve battery drain). However, this modification rendered the monitor's alarm via the nurse call communication port to not function. |
| Product | CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor; CLASSIFICATION NAME: Bed Patient Monitor --- Monitor is connected to a sensor pad that monitors the patient if they attempt to arise from a bed, chair, floor mat, or toilet seat. The monitor will sound an alarm. |
| Quantity | Domestic: 112 units; Foreign: 211 units |
| Distribution | Nationwide, Canada, Ireland, Italy |
| Initiated | 2014-08-05 |
| Classified | 2014-08-26 |
| Terminated | 2017-02-28 |
| Recall number | Z-2554-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Curbell Medical, Inc. (Z-2554-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68961/
Alerts for CURBELL MEDICAL PRODUCTS, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.