Class II — Horiba Instruments, Inc dba Horiba Medical

Recalling firmHORIBA INSTRUMENTS INCORPORATED (DBA) HORIBA Medical
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonHoriba Medical is recalling the ABX PENTRA Magnesium RTU Reagent because the on-board stability claims has been updated. Horiba has updated the ABX PENTRA Magnesium RTU package inserts by removing the 7-day on-board stability claim and replacing it with a 1-day on-board stability claim.
ProductABX PENTRA Magnesium RTU Reagent, Part No. A11A01646 Product Usage: ABX PENTRA Magnesium RTU reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of magnesium in human serum and plasma based on a photometric test using xylidyl blue. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemnia (abnormally low plasma levels of magnesium) and hypermagnesemnia (abnormally high plasma levels of magnesium).
Quantity359 units
DistributionUS Nationwide Distribution.
Initiated2014-08-11
Classified2014-09-23
Terminated2014-11-25
Recall numberZ-2692-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Horiba Instruments, Inc dba Horiba Medical (Z-2692-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/69105/

Not open yet

Alerts for HORIBA INSTRUMENTS INCORPORATED (DBA) HORIBA Medical

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.