Class II — Varian Medical Systems, Inc.
| Recalling firm | Varian Medical Systems, Inc |
|---|---|
| Registered facility | View this company's 14 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Due to user error, the 4DITC can improperly allow users to clear a safety interlock imposed by the 4DITC on behalf of Varian's Bar Code Conical Verification system from the BrainLab's ExacTract console when it is used with BCCV. |
| Product | 4D Integrated Treatment Console (4DITC), versions 10 and 11. 4DITC provided assistance for accurate treatment delivery by monitoring linear accelerator parameters and preventing the radiation therapy device from commencing irradiation when any parameter is out of conformance with the treatment plan. |
| Quantity | 512 affected devices |
| Distribution | Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Brazil, Canada, Chili, China, Columbia, Czech Republic, France, Georgia, Germany, hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Latvia, Lithuania, Macau, Malaysia, Mexico, Netherland,s Portugal, Qutar, Russia, Singapore, Slovenia, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Ukraine … |
| Initiated | 2014-08-18 |
| Classified | 2014-09-16 |
| Terminated | 2015-07-30 |
| Recall number | Z-2654-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Varian Medical Systems, Inc. (Z-2654-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/69126/
Alerts for Varian Medical Systems, Inc
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.